医用防护服阻干态微生物穿透性能测试仪ZGCT0506-A
English name:Medical protective clothing resistance to dry state microbial penetration performance tester
Model:ZGCT0506-A
Product brand:远梓
适用方向:用于测定材料阻抗人体皮屑大小范围内的干态微粒上细菌穿透的性能。
Application direction: It is used to measure the bacterial penetration of dry particles within the size range of human dander.
所属分类:
仪器特点
采用7英寸(320*240)PLC多彩液晶触摸屏,中文菜单显示。采用数字闭环控制,AD采集速度为500KHz,内码100万;高精度时钟芯片,断电存储;PLC控制动态显示加载力值;过载报警停机保护;公称规格、测试、上行、下行、时间、标定,可打印测试数据,,阻干态微生物穿透性能测试仪/干态阻菌仪用于测定材料阻抗人体皮屑大小范围内的干态微粒上细菌穿透的性能,进行测试时,试样被固定在一个容器上。
7 inch (320*240) PLC colorful LCD touch screen, Chinese menu display. Adopts digital closed-loop control, AD acquisition speed is 500KHz, internal code 1 million; High precision clock chip, power off storage; PLC control dynamic display loading force value; Overload alarm stop protection; Nominal specification, test, up, down, time, calibration, printable test data, dry resistance to microbial penetration tester/Dry resistance to bacterial penetration of dry particles within the size of human dander is used to determine the material resistance to bacterial penetration performance on dry particles in the size of the test, the sample is fixed on a container.
仪器参数
完全符合:YY/T0506.5-2009 《病人、医护人员和器械用手术单、手术衣和洁净服——第5部分:阻干态微生物穿透试验方法》中的要求研制,同时符合国际标准ISO 22612:2005 《传染原防护衣 阻干态微生物穿透试验方法》的要求测试。
1、振动形式:气动球式振动器
2、振动频率:20800次/分
3、振动作用力:650N 误差不大于±1%
4、工作台介质:大理石板
5、实验容器:304不锈钢容器
6、工作位:6个
7、工作台尺寸:40cm×40cm×10mm
8、工作电压:220V 50Hz
配备生物安全柜
In full compliance with: YY/T0506.5-2009 Operating sheets, operating gowns and clean clothing for patients, medical personnel and instruments - Part 5: It is developed according to the requirements of the test method for preventing the penetration of dry microorganisms, and meets the requirements of the international standard ISO 22612:2005 "Test method for preventing the penetration of dry microorganisms by infectious agent protective clothing".
1, vibration form: pneumatic ball vibrator
2. Vibration frequency: 20800 times/min
3, vibration force: 650N error is not more than ±1%
4. Workbench medium: marble slab
5. Experimental container: 304 stainless steel container
6, work bits: 6
7, table size: 40cm×40cm×10mm
8. Working voltage: 220V 50Hz
Equipped with biosafety cabinet
售后服务
整机质保壹年
上门培训维修
七天质量问题全款退换货服务
仪器软件终身免费依国家标准规定升级服务、终身维护
The whole machine quality guarantee one year
door-to-door training maintenance
seven days quality problem full-money return service
instrument software life-long free according to the national standard upgrade service, life-long maintenance